41
5−2−4. Zero fine adjustment
(1) Procedures by key operation
Warning ● When the tare weight cancellation (A/Z) and the setting of zero is
executing, and when the zero tracking is effective, the zero fine
adjustment cannot be entered(Displays ER−5). After making the tare
weight cancellation clear(A/Z OFF), the cancellation of the
compensation data(Execution of F−98) and the OFF position of the
zero tracking(Setting “00000” on F−08 and F−09), the zero fine
adjustment mode can be entered.
● During calibration procedures, press the key in case of
interrupting the calibration is required. The calibration data is kept as
they are before entering the calibration and then returns to the
Measurement mode.
● Every time the key is pressed with the load display of “FUNC”,
the display will change as the following arrow marks. Furthermore,
every time the is pressed, the display will change as the reverse
direction of the following arrow marks. However, “VCAL” and “VADJ”
appears only when the optional analog output is attached.
“FUNC”→“CCAL”→“ACAL”→“LCAL”→“ZERO”→“SPAN”→
“TARE”→“CHEK”→“MONT”→“VCAL”→“VADJ”→“FUNC”→
“CCAL”→・・・・・ (Hereinafter, it will repeat.)
Procedures
1
Press the key for about one second.
The load display shows “FUNC”.