What to do if I get a Mechanical Limit Alert on my Misonix sonicone Clinic Medical Equipment?
- RRobert Bowman IIAug 5, 2025
If you get a Mechanical Limit Alert on your Misonix Medical Equipment, it could be due to tip overload. Try releasing the footswitch and reducing tip pressure, or use a higher amplitude setting. Another cause might be a loose or damaged component. In that case, release the footswitch, set the ultrasound to STANDBY, remove the silicone sleeve and probe cover, and inspect the extension probe and ultrasonic tip for damage. Replace if necessary, or re-tighten the extension probe and tip using the correct wrenches. Finally, a defective handpiece could be the reason; if the alert persists after the above steps, the handpiece may need replacement.