What to do if Mizuho osi Hana 6875 Medical Equipment table is dead?
- Oolivia54Aug 2, 2025
If your Mizuho OSI Medical Equipment table is unresponsive, begin by checking the power cord connection to both the AC receptacle and the Motion Control Box. Ensure the power switch is on and illuminated. If it's not lit, the table might be running on battery. Check if the circuit breaker on the Motion Control Box base has tripped; if so, reset it by toggling the power switch off and on. Also, verify the component circuit breakers at the Motion Control Box base haven't tripped; press the buttons to reset them, after identifying the overload source.