EasyManua.ls Logo

Mobius airo2 - Page 16

Mobius airo2
16 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DISTRIBUTOR INFORMATION:
Brainlab AG
Olof-Palme-Str. 9, 81829 Munich, Germany
Europe, Africa, Asia, Australia: +49 89 99 15 68 44
USA and Canada: +1 800 597 5911
Japan: +81 3 3769 6900
Latin America: +55 11 33 55 33 70
France: +33-800-67-60-30
E-mail: support@brainlab.com
MANUFACTURER INFORMATION:
Mobius Imaging, LLC
Phoenix Park, 2 Shaker Road, Bldg F, Suite 100,
Shirley, MA 01464, USA
USA: +1 (978) 796-5068
www.mobiusimaging.com
Email: info@mobiusimaging.com
COPYRIGHT:
This guide contains proprietary information
protected by copyright.
No part of this guide may be reproduced or
translated without the express written
permission of Brainlab.
LIABILITY:
This guide is subject to change without
notice and does not represent a
commitment on the part of Brainlab.
For further information, please refer to
the “Limitations of Liability” section in
the Brainlab Standard Terms and
Conditions of Sale.
Page 16-16
TUBE CURRENT MODULATION (IF REQUIRED)
NOTE: This is only applicable if helical scan mode is
chosen and an A-P scout has been performed.
NOTE: The minimum, maximum and average tube
current values, based on the previous A-P scout is
displayed below
Modulated Tube Current
a.
Select your desired scan mode:
- Constant Tube Current
- Modulated Tube Current
Select
Next
.
FINAL SCANNING
Wait for both marker arrays to be visible a
(indicated by check marks).
Remove the pendant and go to a safe area.
•Press and hold
Scan
s for approx. 15 secs. until
scanning starts and keep pressed for the scan
duration. Follow the display prompts.
Performing 3D Scan
dialog opens when complete.
*60917-32EN*
Document Revision: 1.3
Article Number: 60917-32EN
VERIFY REGISTRATION
Data is sent automatically from
Airo
to
Curve
. Data
can also be sent manually from DICOM Viewer on
the pendant.
Verify registration accuracy by holding the pointer tip
to at least three known anatomical landmarks and
verifying their positions on the screen.
NOTE: Verify registration accuracy in all anatomical
directions (cranial/caudal, medial/lateral, anterior/
posterior).
a
as