2
Molift Air / www.molift.com
General
Product Identification
The Product labels barcode contain product ID,
serial number and production date.
Transport and Operating Conditions
Transport and Storage
For long time storage it is recommended that the
emergency stop button is activated (pulled out).
The lifter can be stored and transported under
temperatures between -25 - 70 °C.
Operating
The lifter is designed for use at standard room
temperatures (+5 to +40°C).
Air Pressure: 70 - 106 kPa
Relative Humidity: 15 - 93 %
Following storage or transport at other tempera-
tures leave the lifter in a room with a suitable
temperature until it reaches a safe operating
temperature.
Medical electrical equipment
requires special precautions regard-
ing electromagnetic compatibility
(EMC). Portable or mobile radio
communication equipment may
affect the medical electrical equip-
ment, and should be kept minimum
25 cm (10 inches) from the lifters
electronics.
Battery and Service indicator
The lifter has two indicator lights.
Wrench symbol / Service light
Battery symbol / Battery light
The electrical system has a power save function
which will turn off the electrical system after ten
minutes without activity. All lights will turn off.
The system is activated when pushing one of the
operating buttons
Battery Light
The battery indicator light will illuminate and
make a sound when battery level is low and the
lifter requires charging. When this occurs the lifter
will have sufficient power available for one full lift
cycle with SWL. When battery is critically low it is
only possible to lower the lifting band.
Service Light
The electronics record the loads lifted and how
much the lift has been used. After a certain period
of operation a signal is given to indicate that
service is required.
Service light Mode
No light Power saving (Stand by)
Green Ready for use
Yellow Order service
Red Perform service
Red + sound Perform service
immediately
Flashing yellow Overheating
Service Scope
Service scope
Service involves replacing the lifting band and
inspection/replacement of worn parts. This must
be carried out by authorised personnel.
Service is needed when the service light is red
(calculated 5.000 lifts in weightclass 2), or service
is required after 5 years if the service counter not
has displayed red yet.
Authorised service personnel can use the Molift
Service Tool to take readings of the lift’s data and
number og lifts.