Priming and Loading the Disposable Set
These instructions are not a substitute for the operator’s manual. Be sure to read the operator’s
manual thoroughly prior to operating the pump. See other side for pump operation instructions.
Quick Reference
Instructions
STEP 1
If you are using the Innity Spike set, remove
protective cover from connector and insert connector
tip into compatible port on the enteral formula
container (image A). Secure the connector by tightly
screwing threaded components together.
If you are using the 500 mL or 1200 mL disposable
set, hold the bag’s funnel upright and pour in feeding
formula. Close cap tightly and hang bag (image B).
STEP 2
Remove protective cover from the feeding
tube connector (image C).
STEP 3 - PRIMING
IF PUMP PRIMING: Loop the silicone tubing around
the pump wheel stretching lightly. Seat cassette
into the pump and close the door (image D).
Turn pump on by pressing and holding the
ON/OFF key until the pump beeps (image E).
Then, press and hold the PRIME key until formula
reaches the feeding tube connector (image F).
IF MANUAL PRIMING: Gently pinch teal colored
tubing below the
symbol (image G). While
pinching, gently squeeze the bag at the same
time to assist uid ow. Hold this position until formula reaches the feeding tube connector.
Loop silicone tubing around pump wheel stretching lightly. Seat
cassette into pump (image D). Close pump door.
FOR BACKPACK USE: Turn the bag upside down and tilt the bag to keep air at tubing port (image H).
Gently pinch teal colored tubing below the
symbol (image G). While pinching, gently squeeze the bag
at the same time to assist uid ow. Hold this position until formula reaches the feeding tube connector.
A B
C D
E F
G H
Speak live with a Moog Clinical Representative for pump questions
and troubleshooting guidance 24 hours per day, 7 days per week.
Clinical & Customer Support 800.970.2337
Visit innityfeedingpump.com for additional information

Other manuals for Moog Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog Infinity and is the answer not in the manual?

Moog Infinity Specifications

General IconGeneral
ModelInfinity
TypeInfusion Pump
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Low battery
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
ManufacturerMoog
Infusion ModesContinuous, Intermittent
DisplayLCD
CompatibilityStandard infusion sets

Related product manuals