Moog CURLIN INFUSION PainSmart - User Manual

Moog CURLIN INFUSION PainSmart
141 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
User Manual
for use with the
PainSmart
and PainSmart
IOD
Ambulatory Infusion System
Moog Medical Devices Group
tf 800.970.2337
ph 801.264.1001
fx 801.264.1051
www.moog.com/medical

Table of Contents

Questions and Answers

  • A
    Amy ChavezSep 12, 2025
    What to do if unable to change amount to be infused to an amount less than what has already infused on Moog Medical Equipment?
    • R
      Rebecca BrownSep 12, 2025
      Recheck data entry. If alert recurs, resume the infusion and observe the Vol Inf / Amt Inf field on the run menu. The value for data entry must be greater than the value displayed on the run menu.
  • J
    Joanna VargasAug 25, 2025
    What to do if unable to change clinician dose while clinician dose in progress on Moog CURLIN INFUSION PainSmart Medical Equipment?
    • C
      Christopher BentleyAug 25, 2025
      If the current clinician dose is to be continued, change the “Clin DOSE” value back to the original value and accept the field and menu. If the current clinician dose is to be terminated, which will allow a new clinician dose to be programmed, enter a value of 0 in the “Clin DOSE” field and accept the field and menu.
  • N
    Nicholas PrinceAug 25, 2025
    What to do if the Moog CURLIN INFUSION PainSmart shows an empty bag?
    • E
      eugenejonesAug 26, 2025
      Verify that the volume remaining in the source container is sufficient to complete the current therapy. Bag volume may have been reduced due to priming.
  • T
    Tyler ChavezSep 23, 2025
    What to do if the library is empty on Moog CURLIN INFUSION PainSmart?
    • T
      Tina MooreSep 23, 2025
      Load a Library or program the pump using the “PROGRAM” option.
  • D
    Diane PhillipsSep 12, 2025
    What to do if a bolus dose is not allowed on Moog CURLIN INFUSION PainSmart Medical Equipment?
    • D
      Donna DavisSep 12, 2025
      If a bolus dose is not programmed and bolus doses are part of the prescription, modify the Rx/medLMTS™ program to allow bolus doses.
  • O
    olivermarkSep 17, 2025
    What to do if bolus is requested when therapy is not PCA on Moog Medical Equipment?
    • T
      tiffanylopezSep 17, 2025
      Remove the remote bolus cord from the pump if connected.
  • D
    Douglas GarciaSep 5, 2025
    What to do if the invalid therapy time is calculated on Moog CURLIN INFUSION PainSmart?
    • M
      Matthew MendozaSep 5, 2025
      Recheck data entry. The changes to the field must comply with the total time limit.
  • J
    jwoodardSep 6, 2025
    What to do if the pump is disabled on Moog CURLIN INFUSION PainSmart?
    • S
      Stephanie McdonaldSep 6, 2025
      The pump needs to be returned for service.
  • A
    Anthony DiazSep 18, 2025
    What to do if number of doses entered exceeds maximum for therapy on Moog CURLIN INFUSION PainSmart Medical Equipment?
    • R
      Ryan PerezSep 18, 2025
      Recheck data entry. The number of doses requested must be less than or equal to the maximum for the therapy.
  • J
    John MillerAug 30, 2025
    What to do if unable to program a field with a value that exceeds the maximum limit on Moog CURLIN INFUSION PainSmart Medical Equipment?
    • C
      carpenteryolandaAug 30, 2025
      Recheck data entry. A value entered cannot exceed the field limit.

Summary

Introduction to the Moog Curlin Infusion Pump

Special Features of the Moog Curlin Infusion Pump

Details the unique features like accuracy, portability, and specialized functions of the pump.

Warnings, Cautions, and Notices

Outlines critical safety information, potential hazards, and important advisories for pump operation.

List of Abbreviations

Provides definitions for abbreviations used throughout the user manual.

Features of the Pump and Administration Sets

Pump Illustrations and Features

Identifies and describes the key components and physical features of the infusion pump.

Disposable Administration Set Illustrations and Features

Details the components and functionality of the disposable administration sets.

Choosing and Installing the Administration Set

Guides users on selecting and properly installing the infusion pump's administration set.

Up Stream Occlusion Alarm Feature

Explains the function and behavior of the pump's upstream occlusion alarm.

Installing the Batteries and Using External Power

Battery Replacement and Door Access

Provides instructions for replacing the pump's C-cell batteries and accessing the compartment.

External Power Sources

Describes options for powering the pump, including C-cell batteries, battery pack, and AC adaptor.

AC Adaptor;Charger

Details the use and connection of the AC adaptor for powering and charging the pump.

Bolus Cord Use

Explains the function and connection of the remote bolus cord for patient-controlled analgesia.

Learning Pump Features

Keypad Function

Details the purpose and operation of each key on the pump's keypad.

Info-On-Demand (IOD) Key Functions

Explains how to access and interpret information displayed via the IOD feature.

Description of the Custom Features

Customized Basal Rate and Bolus Limits

Describes pump configurations for maximum basal rate, clinician, and load dose limits.

Customized Lock Level Setting

Explains the pump's lock level settings and their impact on user access.

LED Light Indicators and Audio Indicator

Details the meaning of LED status lights and the pump's audible alarm system.

Display Screen Overview

Describes the structure and elements of the pump's display screen.

Basics of Programming

Types of Display Screens

Explains the different types of screens encountered during pump programming.

Starting the Pump and System Status Screens

Covers using help, starting the pump, and interpreting status screens like power and maintenance.

Access Code and BioMed Setup Menu

Details the process of entering access codes and navigating the BioMed Setup menu.

Patient History Log and Clearing Data

Describes how to access the patient history log and clear pump data.

PCA Therapy

Prescription Menus and Administration Routes

Explains prescription settings, units, and available administration routes.

PCA medLMTS Setup

Guides on setting medication limitations (medLMTS™) for PCA therapies.

Run Options and Prime Function

Details options available from the Run Options screen and the pump priming process.

PCA Therapy Options

Covers pump customization options like lock settings, clinician dose, and totals tracking.

Starting and Interrupting Infusion

Explains how to start an infusion and pause/interrupt an ongoing therapy.

Resuming and Repeating Therapies

Guides on resuming interrupted therapies and repeating previous prescriptions.

Troubleshooting

Alerts, Alarms, and Error Codes

Introduces the different types of alerts, alarms, and error codes encountered with the pump.

Specific Alarms

Lists and explains common alarm conditions and their resolutions.

Error Codes List

Provides a comprehensive list of error codes, probable causes, and corrective actions.

Accessories

Detachable Pole Clamps

Describes the pole clamps used for mounting the pump or safety shells.

Remote Bolus Cord

Explains the function and use of the remote bolus cord for PCA therapy.

Lockable Safety Shells

Details the protective, lockable cases for securing the pump and accessories.

Pump Holster

Describes the holster for securely mounting the pump onto an IV pole.

Keypad Covers

Lists available protective covers for the pump's keypad.

Cleaning, Environmental Conditions, and Maintenance

Cleaning Procedures

Provides guidelines and prohibited methods for cleaning the infusion pump and its accessories.

Transport, Storage, and Operating Environment

Outlines recommended conditions for transporting, storing, and operating the pump.

Maintenance and Limited Warranty

Details routine maintenance recommendations and the pump's limited warranty terms.

Procedure for Checking Pumps Between Patients

Patient Inspection and Cleaning

Steps for inspecting and cleaning the pump before use with a new patient.

Volumetric Delivery Test

Procedures for performing volumetric accuracy checks on the infusion pump.

Safety System Checks

Steps for verifying the proper function of the pump's safety systems.

Technical Specifications

Pump Specifications Table

Lists physical dimensions, materials, and operating environment specifications.

Power Requirements and Performance

Details battery life, display capabilities, flow rates, and bolus parameters.

Occlusion Detection Times and Volumes

Specifies detection times for occlusion alarms under various conditions.

Clinician Information and Access Code

Clinician Access Code Entry

Explains how clinicians access secured functions using access codes.

PainSmart IOD Catalog Numbers

Catalog Numbers and Configurations

Lists catalog numbers and configurations for different pump models.

Moog CURLIN INFUSION PainSmart Specifications

General IconGeneral
ManufacturerMoog
ModelCURLIN INFUSION PainSmart
IndicationPain Management
Delivery MethodContinuous
Infusion ModesContinuous
Power SourceBattery
Volumetric Accuracy±5%
Flow Rate Range0.1 mL/hr
Volume Range0.1 mL to 9999 mL
Occlusion Pressure SettingsAdjustable
AlarmsOcclusion, low battery, system error
DisplayLCD