EasyManuals Logo

Moog CURLIN INFUSION PainSmart User Manual

Moog CURLIN INFUSION PainSmart
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
72
5. CLR HRLY (Clear Hourly Totals): To clear Hourly Totals press the NO/CHANGE key to
change OFF to ON and then press YES/ENTER. e following screen will appear:
Clear Hourly Totals Screen
a. To clear the Hourly Totals, press the YES/ENTER key. All Hourly Totals will be cleared
and the cursor will return to the Options menu.
b. To maintain the Hourly Totals, press the NO/CHANGE key. Hourly Totals will be
maintained and the cursor will return to the Options menu.
c. A lock level of 2 or 3 will require an access code entry before allowing Hourly Totals reset.
6. medLMTS
(Medication Limits): If this feature was not selected in the Pre-Prescription
Menu, it will not appear on the Options menu. If medLMTS
parameters were programmed
previously, this eld will appear; and if you want to titrate now, select ON and press the YES/
ENTER key to go to the PCA Titrate medLMTS
menu:
PCA Titrate medLMTS™ Screen
is screen allows you to change the basal rate and/or the bolus parameters within the predened
prescription limits. Whenever any value is changed using the medLMTS
option, the new value
will appear on the prescription menu when it is reviewed. If a delta time/rate limitation was
previously programmed, the rate can be increased within those parameters up to 12 times during
the delta time period. If a 13th attempt is made an alert will display: “MAX CHANGES FOR
PERIOD REACHED.
a. BASL RATE: Use the numeric keys to enter the new desired rate. When the correct rate has
been entered, press the YES/ENTER key to accept the rate, and move to the next eld. (see
Figure 5.1).
b. Pt. BOLUS: Use the numeric keys to enter the new patient bolus dose and, when correct,
press YES/ENTER to accept and move to the next eld. (see Figure 5.1.)
c. BOLS INT (Patient Bolus Interval): Use the numeric keys to enter the new patient bolus
time interval and, when correct, press YES/ENTER to accept and move to next eld. e
range for this eld is 0 to 60.
d. #BOLS /hr (number of patient boluses per hour): Use the numeric keys to enter the new
number of patient boluses per hour and, when correct, press YES/ENTER to move to the
nal step in the medLMTS™ menu. e range for this eld is 0 to 15.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog CURLIN INFUSION PainSmart and is the answer not in the manual?

Moog CURLIN INFUSION PainSmart Specifications

General IconGeneral
BrandMoog
ModelCURLIN INFUSION PainSmart
CategoryMedical Equipment
LanguageEnglish

Related product manuals