Figure 2.9 Proper Installation of the Moog Curlin Infusion Administration Set into the pump
Steps in Installing Administration Set into Moog Curlin Infusion Pumps
1. Fully open the door of the pump.
2. Close the slide clamp on the administration set.
3. Activate the Integral Flow-Stop by twisting and removing the breakaway tab.
4. Insert the blue tubing guide into the receptacle on right side of pump in the direction of the
blue arrow, positioning the tubing in front of the door hinge.
5. Center the tubing in the middle of the pumping fingers.
6. Holding the yellow Integral Flow-Stop by the handle (see Figure 2.9), insert it on an angle
into the receptacle on the left side of the pump in the direction of the yellow arrow. CAUTION:
Do not press down on the top of the Integral Flow-Stop when inserting it into the
receptacle. is action could inadvertently open the Integral Flow-Stop and allow uid
to ow through the tubing.
7. Push the tubing slightly into the black Air Detector slot.
8. Open the slide clamp on the administration set.
9. Fully close the door of the pump and latch it securely.
10. If the set is not primed, prime the set according to prime section in the relevant therapy
chapter.
11. Attach the primed set to the patient’s venous access device per agency protocols.
12. Dispose of used administration sets in accordance with all applicable regulatory and
institutional policies and directions.
Figure 1.9 shows the Moog
Curlin Medical pump with
a cut-away of the display
screen, allowing you to
visualize the receptacles
for the small blue Tubing
Guide on the right side of
the pump and the larger
yellow Integral Flow-
Stop on the left side of
the pump. The arrows, in
corresponding colors, are
visible on the background
surface of the pump head
and indicate the proper
direction for placing
these locators into their
receptacles. Also, note that
the breakaway tab has been
removed from the Integral
Flow-Stop, allowing the
administration set to be
placed into the pump
correctly.