51
6. CHANGE LOCK: This selection allows the clinician to change the lock setting of the pump.
is feature is also available under the OPTIONS menu of each therapy. ere are four choices
in this eld: “OFF,” “1,” “2,” or “3.” When the lock is set to OFF, the clinician has access to
all pump functions. Each level higher limits the options available to the patient or caregiver
(see each lock option screen in the individual therapy sections and/or the lock level table in the
chapter entitled “Clinician Information and Access Code”). If the desired lock does not appear
on the display, press the NO/CHANGE key until the correct lock does appear. en, press the
YES/ENTER key to accept and move the cursor to the next eld.
7. MANUAL PROGRAM: This field is used to customize the Moog Curlin Infusion pump to
ENABLE/DISABLE the PROGRAM menu option on the main menu. ere are two choices:
“OFF” and “ON.” When MANUAL PROGRAM is set to “OFF” the PROGRAM menu
option is not available on the main menu.
a. To enable MANUAL PROGRAM, press the NO/CHANGE key until the selection states
ON.
b. To disable MANUAL PROGRAM, press the NO/CHANGE key until the selection states
OFF.
c. When the ON/OFF status is satisfactory, move the cursor to the last eld, “ACCEPT?” and
press the YES/ENTER key to accept. To abort the enable therapy process, press the NO key
to exit this menu and return to the BioMed Setup Menu.
8. ENABLE THERAPY: This field is used to customize the Moog Curlin Infusion pump to the
specic therapy ordered for the patient. When you select this function by pressing the YES/
ENTER key, the menu below appears, listing all ve of the possible therapies. e pump’s
default setting is to enable PCA and disable all other therapies.
Enable Therapy Menu
You will not be allowed to change the default setting. If you attempt to enable a therapy, the
following prompt will appear of this fact.
Basics Programming