PCA medLMTS™ Setup Screen
a. MAX RATE (Maximum Rate) limits: Use the numeric keys to enter the upper limits of the
infusion rate. When the correct maximum medLIMIT rate is entered, press YES/ENTER and
the cursor will move to the next eld. e minimum rate is 0, and the maximum rate must be
equal to or less than maximum Basal Rate listed in Figure 5.1 Administration Routes table.
b. ∆Rate (Delta Rate): Some prescriptions include incremental levels or delta limitations (e.g.,
“Do not raise the rate more than 1 mL in 12 hours”). If the prescription has these components,
use the ∆Rate eld and the ∆ Time eld to enter those limits. Example: Enter the incremental
medLIMIT amount level of 1 mL on this eld, and press YES/ENTER to move to the next eld.
c. ∆Time (Delta Time): Enter the incremental or delta time limit. (Example: enter 12:00
hours.) If there are no incremental or delta level components to the prescription, leave these
elds (∆Rate and ∆Time) set at their default settings of “0” and press YES/ENTER to move to
the next eld.
d. Max BOLS (Maximum Bolus Dose): Use this field to set the upper limit of the Patient Bolus
or Patient Demand dose. If there is no bolus dose, enter “0.” If a value is given for the Bolus
Dose, use the numeric keys to enter it on this eld, press the YES/ENTER key to accept, and
move to the next eld. e value entered for this eld must be equal to or less than maximum
Bolus Dose listed in Figure 5.1 Administration Routes table.
e. MinINT (Minimum Bolus Interval): Use this field to set the minimum amount of time
allowed between patient bolus doses. When the correct time is entered, press the YES/ENTER
key to accept, and move to the next eld.
f. #BOLS/hr (Number of Boluses allowed per hour): Use this field to enter the maximum patient
bolus/hr. When correct, press the YES/ENTER key to accept, and move to the next eld.
g. ACCEPT? If there are any changes to be made, use the up or down arrows to move to that
eld, use the numeric keys to enter the corrected data, and press YES/ENTER to accept the new
values. Move the cursor back to the ACCEPT? eld and press
PCA Therapy