74
If you press the NO key to indicate a lter is not in use, the Air-In-Line sensor will be set to
detect the presence of any contiguous 2-milliliter air in the line. e alert message below will
then briey appear on the screen. If the alert occurred at the Run Options screen, the pump will
start the infusion.
Pump Will Sense 2mL Alert Screen
Also, when the Air-In-Line sensor is completely disabled and the infusion goes to an “Infusion
Complete” or “Empty Bag” status, the Air Sensor setting will automatically change to 2 mL.
9. AUDIO (Audio Volume): This feature allows you to adjust the audio volume of the pumps
alarm setting. You can enter a number from 9 (the loudest setting) to 1 (the quietest setting).
When you select a number, the pump will beep at that audio level. Press the NO key and enter
a new setting (1 – 9). Repeat until the desired setting is reached, and then press YES/ENTER
to select and move the cursor to the next eld.
10. PWR CK? (Power Check): This feature allows you to periodically recheck the battery charge.
It will display the active power source of the pump, whether internal batteries or external power.
If the active source is the batteries, a graph will indicate approximately how much charge
remains in the batteries. To access this information, select ON and press YES/ENTER. e
Power Source Screen will display for ve seconds and return to the Options Menu.
11. ACCEPT OPT? If the changes made in the Options Menu are acceptable, press the YES/
ENTER key. If the changes are not acceptable, cursor up to any eld for corrections or abort all
changes made in this menu by pressing the NO key, and return to the previous screen.
There are two remaining selections to consider from the Run Options Screen: “NO to REVIEW
and “RUN to START.”
Reviewing or Changing the Program
To review or make any changes in the prescription programming, press the NO/CHANGE key at
the Run Options screen, and the prescription screens will reappear (changes may be inhibited by
lock setting). Changes to the prescription can be made before the start of the therapy; however,
once a therapy has started, no changes can be made to the Pre-Prescription Screen. It will appear
only for review purposes. Attempts to change the Pre-Prescription Menu during a new program
review will cause an alert screen to be displayed, notifying the programmer that changes will require
reprogramming of the prescription.

Table of Contents

Questions and Answers: