77
Hanging a New Container
2. REPEAT Rx: To repeat the present therapy and add another IV container with the same
prescription, press the YES/ENTER key at this eld. If more than one therapy is programmed,
a Select erapy Screen will list all preprogrammed therapies. Select which therapy to repeat by
moving the cursor to the therapy of choice and pressing the YES/ENTER key.
e pump will then display the program screen with the last programmed values of the
therapy and the cursor will automatically advance through the prescription menus. You can
stop the auto-advance function by pressing any key (except ON/OFF) and then complete the
conrmation of the prescription manually by pressing the YES” key at the remaining elds.
Note: e previously programmed AIL setting is also retained
When the program is accepted, the display returns to the above “Run Options Screen.” If the
current therapy is still in progress when the “REPEAT Rx” function is selected, the following
screen will appear:
a. is screen is a reminder that the infusion is not complete. If the incomplete therapy is
to be continued, press NO/CHANGE and the Pause Menu screen will be displayed, then
press YES/ENTER and the Run Options screen will be displayed. To resume the unnished
therapy, press RUN.
b. If the therapy is to be terminated and a new IV container added, press YES/ENTER to
repeat the infusion. e following screen will then appear, asking whether or not you wish
to maintain the current “Bolus or MedLMTS
schedule.
Press YES/ENTER to maintain the current schedules and continue with the programming of
the pump. Press NO/CHANGE to start a new schedule. e following screen will then appear
asking you to conrm the clearing of the schedules.
Press YES/ENTER to clear the current schedules and continue with the programming of the
pump. Pressing NO/CHANGE will return you to the prior screen.
PCA Therapy