Shift Totals Screens
a. Cleared/ELAPSED: is eld displays the date and time the shift total was last cleared. It
toggles with the amount of time elapsed since that clearing.
b. Vol INF/Amt INF (Volume/Amount Infused): is eld displays the volume (mL) and
amount (mg or mcg) infused since the shift total was cleared.
c. BOLUS/Clin DOSE (Clinician Dose): is eld displays the number of clinician doses
given since the total was cleared. It toggles with the number of patient activated boluses
received of those requested.
d. To exit the shift total feature without clearing the totals, press the YES key at the
highlighted EXIT eld.
e. To clear the Shift Totals, press the NO key at the “EXIT” eld. e eld will change to read
“CLEAR,” press the “YES” key now to clear. A lock level 3 will require an access code entry
before allowing the user to clear the Shift Totals.
4 HRLY TOT (Hourly Totals): To view hourly infusion totals press the NO/CHANGE key
to change OFF to ON and then press YES/ENTER. Either of the following two screens will
appear:
Hourly Totals Screen
or
e screen displays hourly infusion delivery for each hour (00-23) of the last 24 hours of
therapy in the unit selected.
a. e highlighted eld has a marker
indicating the current hour in progress.
b. “Volume” or “Amount” will show in the rst column, “BG” (Bolus Given) in the second
column and “BA” (Bolus Attempted) in the third column.
c. Use the up arrow key to scroll back through the last 24 hours.
d. If the infusion was not running, a series of ------ will appear for that hour.
e. To exit this screen, press the NO/CHANGE key.