42
Description of the Custom Features
A customized Moog Curlin Infusion PainSmart™ IOD pump may have Maximum Basal Rate
Limit, Maximum Pt. Bolus Dose, Clinician Dose, and Load Dose Limits values that are different
from default maximum values for all administrative routes (IV, EPI, and SQ), and a different Lock
Level other than the default value OFF. These values, once programmed at the factory, can not be
changed unless the pump is initialized or a new setting is programmed, both of which must be
performed by Moog Curlin or factory trained BioMedical professionals.
Customized Maximum Basal Rate Limit
A customized Moog Curlin Infusion PainSmart
™ IOD pump may come with a Maximum Basal
Rate Limit value set in the factory that is different from the default maximum value.
An alert will be displayed if a value greater than the factory set Maximum Basal Rate Limit value is
entered in BASL RATE, MedLMTS Mx Rate, and MedLMTS Delta Rate.
For example, while programming a PCA (IV) therapy in a customized Moog Curlin Infusion
PainSmart
™ IOD pump, where the Maximum Basal Rate Limit is factory set to be 9.9 for IV
administration route, if you enter 21 for BASL RATE, an alert message will appear:
“GREATER THAN MAX. FIELD LIMIT MAX=9.9”
The values you are allowed to enter in this example, for BASL RATE, are between 0.1 and 9.9, with
an increment of 0.1. Same limit will apply to MedLMTS Mx Rate, and MedLMTS Delta Rate also,
if the therapy calls for them.
Customized Maximum Bolus, Clinician, and Load Dose Limits
A customized Moog Curlin Infusion PainSmart
™ IOD pump may come with Maximum Pt.
Bolus, Clinician, and Load Dose Limit values set in the factory.
An alert will be displayed if a value greater than the factory set value is entered in Load Dose,
Clinician Dose, Pt. Bolus, and MedLMTS Pt. Bolus.
For example, while programming a PCA (IV) therapy in a customized Moog Curlin Medical
PainSmart
™IOD pump, where the Maximum Pt. Bolus Dose Limit is factory set to be 9.9 for IV
administration route, if you enter 10 for LOAD DOSE, an alert message will appear:
“GREATER THAN MAX. FIELD LIMIT MAX=9.9”
The values you are allowed to enter in this example, for LOAD DOSE, are between 0.1 and 9.9,
with an increment of 0.1. Custom limits also apply to Clinician Dose, Pt. Bolus, and MedLMTS
Pt. Bolus also, if the therapy calls for them.
Learning Pump