52
Patient History Log
e patient history log is a cumulative record of the approximately last 6,000 events that have
occurred in the pump. Example: When prescription information is saved using the YES/ENTER
key at the “DONE” or “ACCEPT” elds, a copy of the information also gets date and time
stampedand is recorded in the patient history log. Alarms, error codes, certain alerts, and certain
key press events also get logged when they occur. e patient history log can be printed using
the Moog Curlin Infusion print package and by selecting the “Print Pt. Hx” option. Selecting
the “Clear Hx Log” option will erase the patient history log. When the history log is full, it will
overwrite on a rst-in, rst-out basis.
Note: The patient history log should be printed before clearing occurs if a hard copy record is
required. e history log, when printed, lists the most recent event rst. Clearing the history
log does not aect the saved patient information.
When printed, the patient history log includes a heading at its beginning and a footer at its end.
The heading includes the pump serial number, the patient history log title and revision number
information of the software and hardware of the pump. The footer contains information about
how much of the history log has been used. The information is printed with the most recent event
first, and organized for easy transfer to a spreadsheet for sorting and reporting. A partial sample of
a history log is given on the following page: