54
When you select “YES,” the display will read “Printing Patient History Record.”  When 
the printing is completed, the pump display returns to the BioMed Setup Menu. To abort 
the printing process, press the NO/CHANGE key on the pump’s keypad, and the display 
will return to the BioMed Setup Menu. If the Patient History Log stops before all data is 
transferred, exit and restart the “Print Pt. Hx” command.
Note: To utilize the print feature of the Moog Curlin Infusion pump, you must have a PC 
running the hyperterminal program within MS Windows
™, and use the Moog Curlin 
Infusion Data Interface Cable which is purchased separately. Detailed user instructions are 
provided with the cable.
Clearing the History Log
10.  CLEAR Hx LOG: This feature is used to clear the patient history log from the pump. When 
   you select YES/ENTER, the pump displays:
 When you select “YES”, the history log is cleared and the display returns to the BioMed Setup 
Menu. Selecting “NO” aborts the feature and also returns to the menu.
Clearing Patient Information 
11.  CLEAR Rx: When this command is selected, prescription information for all currently
 programmed therapies is cleared; however, the patient data is retained. When you select YES/
ENTER the pump displays:
 
 When you select “YES”, all prescription programs are cleared, and the display returns to the 
BioMed Setup Menu. Selecting “NO” aborts the feature and also returns to the menu.
12.  CLEAR Pt. DATA/RX: Selecting this feature clears all data programmed. The history log 
   is NOT cleared. When you select YES/ENTER at this eld, the following screen displays:
When you select “YES,” all patient data clears and the display returns to the BioMed Setup 
Menu. Selecting “NO” aborts the feature and also returns to the menu.