37
Keypad Function
1. ON/OFF: Used to turn the pump on and off. When the pump is running, however, the
PAUSE key must be used to stop the infusion rst and then the OFF key can be pressed to
turn the pump o. is requirement prevents the interruption of an infusion by a single
accidental press of the OFF key.
2. RUN/PAUSE: Used to start or pause the motor of the pump, which, in turn, starts or pauses
uid delivery. When the pump is running, pressing this key will pause the pump. When the
pump is paused, pressing this key during the Run Options menu will start the pump. is key
is also used to terminate a pump alarm.
Note: If the cause of the alarm is not resolved when the pump is resumed, the alarm will
recur.
3.
Up arrow: Used to scroll the highlight bar or field cursor up on the display screen.
Note: e Up arrow will not accept a selection which must be made with a YES/ENTER or
NO/CHANGE key.
4.
Down arrow: Used to scroll the highlight bar or field cursor down on the display screen.
Note: e Down arrow will not accept a selection which must be made with a YES/ENTER or
NO/CHANGE key.
5. NO/CHANGE: Used to respond “NO” to data or questions presented or to reject or to
change highlighted data.
Note: e Moog Curlin Infusion pump requires you to actively change a selection. e Up
arrow or Down arrow will not change a selection.
6. YES/ENTER: Used to respond YES/ENTER to data or questions presented or to accept
highlighted data when prompted to press ENTER key.
Note: e Moog Curlin Infusion pump requires you to armatively accept a selection. e
Up arrow or Down arrow will not accept a selection.
7. HELP/OPTIONS: Used to request a help screen when the pump is paused or is at the
“Infusion Complete Screen.” It is also used to access the Options Screen when the pump is
running. When the Options Screen is displayed, help is again available with this key. When in
a help screen, use the arrow keys to scroll through the help message, and then press the HELP
key again to exit Help.
8. ./SILENCE: Used as a decimal point when programming numeric data. Used also to
temporarily silence the alarm (for one-minute intervals).
9. PRIME/BOLUS: Used to prime fluid through the tubing to clear it of air or to administer a
patient demand bolus dose of medication in PCA therapy.
10.0 - 9 Numeric keys: - Used for data entry and the display of Info-On-Demand (optional)
functions when the pump is in the RUN mode.