76
PCA Information Screens
Interrupting an Infusion
The infusion can be paused at any time by pressing the PAUSE key. The yellow STANDBY LED
light blinks, and the following screen displays:
Pause Menu Screen
The screen offers the user three options:
Resuming the Current Container
1. RESUME: To resume the therapy in progress from where it left off, press the YES/ENTER key
at this field. The pump will return to the Run Options Screen, and you can resume the therapy.
Note: If you wish to hang a new container, move the cursor to REPEAT Rx (see following
section).
Run Options Screen
a. Press the RUN key to resume the therapy.
b. Press the NO key to go back to the programming screen to make changes in the therapy
program (this feature may be blocked by a security lock, thus not allowing changes to the
program unless an access code is given). Also, remember that no changes are allowed in the
rst prescription menu if an infusion is in progress.
c. Press the OPTIONS or PRIME keys if either of these functions is desired at this time.
Note: e top line of these running screens will display the name of the running protocol when
programmed in the Library mode, or it will display "Unidentied" when programmed outside of
the library.