57
Resuming a Therapy
1. RESUME—If a therapy is interrupted and the pump is turned off or paused before completion
of the infusion, you may select “RESUME” by moving the cursor to this eld and pressing
the YES/ENTER key. A notication screen displays, and a prompt is given to run the pump.
At this time, you can continue to infuse the prescription from the exact point at which it was
interrupted.
Note: Patients should be taught proper time frames or limitations for leaving their IV bags or
medication reservoirs hanging during an interrupted state and, if indicated, to change
to a new bag according to the healthcare provider’s policies and drug manufacturer’s
instructions.
Repeating a Pre-Programmed Therapy
2. REPEAT Rx—Use this feature if an infusion is complete and you wish to repeat the same
prescription over again or if a therapy is programmed and the pump is turned o before the
infusion is started. Select this eld by moving the cursor to REPEAT Rx, and press the YES/
ENTER key.
If the pump has been programmed for more than one therapy and you wish to select a therapy
at this time, move the cursor to “REPEAT Rx,” and press the YES/ENTER key. is action will
take you to the Select erapy Screen and allow you to choose which of the programs you now
wish to repeat.
Note: If you are choosing a therapy that is different than the last one selected, Shift Totals and
Hourly Totals will be cleared.
If only one therapy is programmed and “REPEAT Rx” is selected, the cursor will automatically
advance through the prescription menus. You can stop the auto-advance function by pressing
any key (except ON/OFF) and then complete the conrmation of the prescription by pressing
the YES key at the remaining elds.
Note: REPEAT Rx does not allow any changes to the Pre-Prescription Menu. If changes need
to be made in Pre-Prescription selections (units, concentration, etc.) select “New Program
and reenter the entire prescription.
e lock settings aect the use of “REPEAT Rx.” Lock 1 allows REPEAT with changes to the
prescription screen only. Lock 2 allows REPEAT but does NOT allow the user to make changes
in the prescription. Lock 3 does not allow REPEAT. (See lock setting table.)
If a therapy has not completed and the “REPEAT Rx” function is selected, the following screen
will display: