39
Info-On-Demand (IOD) Key Functions
Designation and functionality of the ten numeric keys of the Info-On-Demand keyboard:
1. Date and Time (DateTime): The Date and Time screen displays the current date and time
according to the pump. Changes to this screen are not allowed.
2. Power Check (PowrChck): The Power Check screen displays the active power source of
the pump, whether internal (batteries) or external. If the active source is the C-cell batteries, a
bar graph will indicate the amount of power remaining in the batteries. is graph is only an
approximation of the charge left in the batteries and not intended to give specic representation
of the battery life. Furthermore, the bar graph should NOT be used to check the state of the
external NiMH battery pack, because it will not provide with accurate results.
3. Option Info (OptnInfo): This screen provides information on selected options. The
information displayed here includes the status of the Lock Level, Down Occlusion, Air Sensors,
and the Audio Level. Changes to this screen are not allowed.
4. Hold Screen (HoldScrn): Pressing the Hold Scrn/4 key will hold the current Run screen
information for an additional 8 seconds and then the pump will return to the normal running
screen. Changes to this screen are not allowed.
1
Date
Time
2
Powr
Chck
3
Optn
Info
4
Hold
Scrn