8. Percent Bolus (%Bols): is screen provides Bolus Volume as percentage of Total
Volume. TBOL is the total volume infused through bolus, TVOL is the total volume
infused, and % BOLS is the percentage of TBOL over TVOL. Changes to this screen are
not allowed.
9 . Clear Shift Total (CLRShft): Two Shift Total screens are displayed for 11 seconds
when this key is pressed. is key allows the user to view or CLEAR the Shift Totals (this
action does not interfere with the operation of the pump).
To CLEAR the Shift Totals information to zero, press the CLRShft/9 key, and then press
the YES/ENTER button in response to the “DONE? CLEAR” question and the pump
will go back to the running screen. Press the CLRShft/9 key again to verify that the values
are reset or cleared.
To EXIT the Shift Totals screens without clearing, press NO/CHANGE for the “DONE?
CLEAR” question and then YES/ENTER for “DONE? EXIT” and return to the running
screen. e pump will return to the running screen automatically after 11 seconds have
elapsed since the pressing of the CLRShft/9 key, even if no key is pressed.
10. Check Maximum Infusion (ChckMax): e information displayed here is the
maximum (rate or amount) that can be infused hypothetically, given the prescription
parameters. is “Maximum Infusion per Hour” serves as a “sanity check” after the pump
has begun infusing. Changes to this screen are not allowed.
%