EasyManua.ls Logo

Moore Industries RIY - Page 11

Moore Industries RIY
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIY
The Interface Solution Experts
9
DECADE
RESISTANCE
BOX
RIY
1
2
3
4
+PS
–PS
250
+
DC VOLTMETER
12-42 VDC
POWER
SOURCE
+
NOTES:
1. Terminals 1 and 4 are the primary connection terminals.
2. To simulate a 3-wire input, use terminal 3 for lead length
compensation
3. To simulate a 4-wire input, use terminals 2 and 3 for lead
length compensation.
(SEE NOTES)
Figure 4 illustrates the calibration setup required to
perform standard ranging.
To monitor the output, a dc voltmeter and a precision
load resistor are called for. Volmeter readings of 1-5
Vdc read across a 250 ohm resistor represent 4-20
mA, proportionally. An output reading of 3 Vdc is
equal to a 12 mA output with a 250 Ohm resistor.
Similar calculations can be made over the 1-5 Vdc
range with the following conversion formula: V
(voltage reading)/250 = mA.
Table 3. RIY Calibration Equipment (Standard Ranging)
Equipment
Decade Resistance
Box
Voltmeter and
Precision Resistor
DC Power Source
Specifications
Accuracy of ±0.01%, or better
Digital voltmeter, accuracy of ±0.005% or better; 250 precision
resistor, tolerance of ±0.01%
12-42 Vdc
Standard Ranging Procedure
1. Set range and other configuration switches for the
required RTD or ohms input, as necessary (refer
to figure 2 and table 2).
NOTE
Switch SW301-5 must be set to the ‘off’
position to perform standard ranging,
and set switch SW301-6 to the ‘off’
position for units with this switch
functional for keyboard enable.
Figure 4. Calibration Hookup Diagram (Standard Ranging)

Related product manuals