EasyManua.ls Logo

MRC TES - 3060 - Audible Continuity

MRC TES - 3060
31 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
Peak indication
Range Resolution Accuracy Overload protection
20A~80A 0.1A
±
(10% + 10)
1100A
80A~1000A 0.1A
±
( 6% + 10)
1100A
Peak detect acquisitions time
1ms
Peak indication
Range Resolution Accuracy Overload protection
20V~80V 0.1V
±
(10% + 10)
600Vrms
80V~600V 0.1V
±
( 6% + 10)
600Vrms
Peak detect acquisitions time
1ms
Audible continuity
Range
Continuity
beeper
Open circuit voltage
Overload
protection
< 0.050
3.2V
600Vrms
Resistance (
Ω
)
Range Resolution Accuracy
Open circuit
voltage
Overload
protection
10K
Ω
1
Ω
±
( 1
+ 5)
3.2 V
600Vrms
Diode test
Range Resolution Overload protection
0.001V 600Vrms
Frequency ( Hz )
Range Resolution Accuracy
Voltage
sensitivity
Overload
protection
1KHz 0.1Hz
5KHz 1Hz
±
(0.5% + 5)
10V or 10A 600V / 1100A
Optional accessories:
RS-232 interface adaptor / software