Maintenance and Cleaning
SavOx / SavOx Industry
18
GB
4.3 Tightness Test
The tightness test of the device is carried out using the MSA tightness tester for
chemical oxygen breathing protection apparatus. Detailed descriptions of the struc-
ture, function, use and maintenance of the tightness tester can be taken from the
relevant operating manual of the tightness tester [ chapter 8].
The tightness test of the device is carried out as follows:
(1) Push the device without any straps into the tightness tester in such a way that
the cover points towards the door.
(2) Close the door and lock.
(3) Set timer to 90 s.
(4) Create overpressure with the lever, i.e. push the lever downwards until it
engages and start the timer at the same time.
(5) After 30 seconds, cover the black pointer with the red pointer.
(6) After a further 60 seconds, the timer signals the end of the test period. Then
read the pressure difference between the red and the black pointers.
The device is tight if the black pointer remains in the prescribed pressure range and
if the pressure difference is no more than 5 mbar.
If the pressure drop is more than 5 mbar, the unit must be replaced.
4.4 Indicator - applies only for SavOx
Perform an indicator check to controll if the device must be replaced. The indicator
is a circle with two semi-circles, the left side is the indicator, on the right side a com-
parison colour is shown.
A blue colour on the left side indicates a functionable device. A pink colour or col-
ourless indicator on the left side indicates, the device must be replaced. To verify
the colours, a comparison colour in pink on the right side is shown.
(1) Check the indicator on the left side.
(2) Compare the left side with the right side.
The colour of the left side must be blue.
Otherwise replace the device.
Blue / Pink Pink / Pink