2 Unit Connection
! On test stand left side, ensure that all 5 barb ttings are in
place & straight before connecting cylinder holder or
another test stand.
" Press 2 units together until barb ttings are fully inserted &
ange screw holes align.
# Insert 3 screws: to connect multiple test stands, remove
white gas plugs from all units except farthest right test stand.
Continue adding test stands to right, cylinder holders to left.*
GALAXY® GX2 Automated Test System Quick Start Guide -
Description and Setup
!
"
$
#
%
&
Memory card port provided on test stand is for calibration &
bump record storage; SD or SDHC memory card can be
purchased from MSA or elsewhere.
Test stand Ethernet jacks
1 System Components
! Touch screen display
" Status LED
# Test stand to test stand connector
% SD card port
$ USB port
& Gas plugs
If using MSA cylinder holders, left-most cylinder holder should
have fresh air lter attached to top port.
*If using ammonia or chlorine test gas, read restriction found under instruction manual chapter 2.8 “Special Conditions for Use with
Reactive Gases” .
GB/EN