Light Meter. User Manual
Warnings
Explosion hazard. Do not use this device in the presence of

Read and be familiar with this instruction manual before
using this device.
Only use this device under the direct supervision of a
licensed medical practitioner.
Inspect this device before each use to ensure proper
functioning.
!
!
!
!
The hospital/facility is responsible for ensuring that all per-
sonnel who operate or maintain this device are trained in its
operation and safe use, and for maintaining training records
of attendance and evidence of understanding.
Operation
Taking Measurements - Overhead Phototherapy
1. Turn the MTTS Light Meter on.
2. Hold the MTTS Light Meter against the infant’s body as near to the
umbilical as possible and aim the Light Meter at the center of the

- For reproducible measurements, always hold the sensor on the
same place on the infant’s body.
- Changes in the distance or angle of the light to the patient will
change the irradiance the patient receives, requiring new measure-
ments to be taken.

touch panel, do not press hard
Phototherapy light
Sensor aimed for
maximum reading