Sequence of button presses to set PF to 200 and TR A to 0, indicating readiness for calibration.
Rinsing and filling the well with Distilled Water, then initiating calibration.
Rinsing and filling the well with S0920 standard, pressing READ/ENT for calibration.
Checking if the calibration reading is within the 190-210 ppm range.
Steps to test the pen fluorometer calibration using S0920 solution.
Successful calibration shows readings between 190 and 210 ppm.
Collecting, preparing, and measuring chill water samples with the fluorometer.
Multiplying the fluorometer reading by 40 to get TRAC107 ppm.