5 of 17
Copyright 2014 ▪ Nanalysis Corporation
• Located in a temperature stable environment of 18-26
o
C.
Equipment will not operate properly if the temperature is outside
of this range.
• Temperature variation limits. Either:
1. Temperature variations no greater than +/- 1
o
C from nominal,
regardless of rate of change (Controlled temperature environment)
2. Temperature variations no greater than +/- 3
o
C from nominal,
having a rate of change not exceeding 0.5
o
C per hour. (Uncontrolled
or loosely-controlled temperature environment)
• Should not be situated near doors that are in constant use
(open/close) or air conditioning vents. Cold or warm air blowing on
the instrument will affect its performance.
• Should not be within 2.5 meters of electric motors or large masses
of moving metal.
• If situated on a desk or bench it should not be over the drawer
stack, as moving steel will affect magnetic gradients.
Site Requirements