Located in a temperature stable environment of 18-26
o
C.
Equipment will not operate properly if the temperature is outside
of this range.
Temperature variation limits. Either:
1. Temperature variations no greater than +/- 1
o
C from nominal,
regardless of rate of change (Controlled temperature environment)
2. Temperature variations no greater than +/- 3
o
C from nominal,
having a rate of change not exceeding 0.5
o
C per hour. (Uncontrolled
or loosely-controlled temperature environment)
Should not be situated near doors that are in constant use
(open/close) or air conditioning vents. Cold or warm air blowing on
the instrument will affect its performance.
Should not be within 2.5 meters of electric motors or large masses
of moving metal.
If situated on a desk or bench it should not be over the drawer
stack, as moving steel will affect magnetic gradients.