2 - User interface
Main screen
1. START frequency 2. STOP frequency
The START frequency and STOP frequency are shown at the bottom of the display.
3. Marker
The marker position for each trace is displayed as a small numbered triangle. The selected marker can be moved to
any of the measured points in the following ways:
Drag a marker on the touch panel – best to use a stylus for this.
Press and hold the JOG LEFT or JOG RIGHT buttons.
4. Calibration status
Displays the saved slot number of the calibration being used and the error correction applied.
C0 C1 C2 C3 C4 : Each indicates that the corresponding calibration data is loaded.
D : Indicates that port 1 3-term error model is applied.
5. Reference position
Indicates the reference position of the corresponding trace. You can change the position with:
DISPLAY →SCALE →REFERENCE POSITION.
6. Marker status
The active marker that is selected and one marker that was previously active are displayed top right.
7. Trace status
The status of each trace format and the value corresponding to the active marker are displayed.
For example, if the display is showing: CH0 LOGMAG 10dB/ 0.02dB , read it as follows:
Channel CH0 (reflection)
Format LOGMAG
Scale is 10dB
Current value is 0.02dB
For active traces, the channel name is highlighted.
8. Battery status