CALIBRATION PROCEDURE
NI 6601/6602
This document contains information and instructions for calibrating the
National Instruments 6601/6602 data acquisition devices.
Contents
Introduction............................................................................................. 2
What Is Calibration? ........................................................................ 2
Why Should You Verify? ................................................................ 2
How Often Should You Verify? ...................................................... 2
Software and Documentation.................................................................. 2
Software.................................................................................... 3
Documentation.......................................................................... 3
Test Equipment ....................................................................................... 3
Test Conditions ....................................................................................... 4
Calibration Process ................................................................................. 4
Setting Up the Stable Clock............................................................. 5
Setting Up a PXI-6608 Clock Source....................................... 5
Setting Up an Alternative Clock Source................................... 8
Connecting the Clock Source to the NI 6601/6602 Device............. 8
Connecting a PXI-6608 to the NI 6601/6602 Device............... 8
Connecting an Alternative Clock Source to the
NI 6601/6602 Device............................................................. 9
Measuring the Frequency of the NI 6601/6602 Device................... 9
Comparing the Measured Frequency to the NI 6601/6602
Device Specifications.................................................................... 17
ni.com/manuals
DeutschFrançais

Other manuals for National Instruments NI 6608

Question and Answer IconNeed help?

Do you have a question about the National Instruments NI 6608 and is the answer not in the manual?

National Instruments NI 6608 Specifications

General IconGeneral
Form FactorPCI
Maximum Output Frequency10 MHz
Maximum Source Frequency20 MHz
Timing Resolution50 ns
Counter Resolution32 bits
Input Voltage Range0 to 5 V
Timebase Accuracy50 ppm
Connector Type68-pin VHDCI
Bus TypePCI
Number of Channels/Counters/Timers8
Output Voltage Range0 to 5 V
Power Requirements5 V
Operating Temperature Range0 °C to 55 °C