EasyManuals Logo

National Instruments PXIe-5413 User Manual

National Instruments PXIe-5413
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
Table 5. AC Amplitude Accuracy Verification
DMM Range
(V
RMS
)
Amplitude Range
(V
pk-pk
)
Expected
Amplitude
(V
pk-pk
)
As-Found
Test Limit
(mV)
As-Left Test
Limit (mV)
1 5 7.34 7.34 ±74.40 ±32.66
2 2.27 2.27 ±23.70 ±10.10
3 2.18 2.18 ±22.80 ±9.70
4 2.1 2.1 ±22.00 ±9.35
5 2.02 2.02 ±21.20 ±8.99
6 1.94 1.94 ±20.40 ±8.63
7 1.85 1.85 ±19.50 ±8.23
8 1.77 1.77 ±18.70 ±7.88
9 1.68 1.68 ±17.80 ±7.48
10 0.5 0.69 0.69 ±7.90 ±3.07
11 0.21 0.21 ±3.10 ±1.50
12 0.05 0.065 0.065 ±1.65 ±0.81
1. Connect one end of the SMA (m)-to-BNC (m) cable to CH 0 of the PXIe-5413. Connect
the other end to the double banana plug-to-BNC (f) adapter, and connect the double
banana plug-to-BNC (f) adapter to the HI and LO connectors of the DMM.
2. Configure the DMM with the following settings:
Function: AC Voltage
Resolution: 6.5 digits
Aperture time: 100 ms
Number of averages: 4
Minimum frequency: 49 kHz
Auto zero: Enabled
ADC calibration: Enabled
Vertical range: DMM range listed in the AC Amplitude Accuracy Verification table
3. Configure the PXIe-5413 with the following settings:
Channel: Channel of the PXIe-5413 connected to the DMM
Load impedance: 1 MΩ
Output mode: Standard Function
Waveform type: Sine
Frequency: 50 kHz
10 | ni.com | PXIe-5413 Calibration Procedure

Other manuals for National Instruments PXIe-5413

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXIe-5413 and is the answer not in the manual?

National Instruments PXIe-5413 Specifications

General IconGeneral
BrandNational Instruments
ModelPXIe-5413
CategoryInverter
LanguageEnglish