EasyManuals Logo

NDD Easy on-PC User Manual

NDD Easy on-PC
2 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #2 background imageLoading...
Page #2 background image
Doc 0327 iss02
8
10
9
11
Perform the test. When the manoeuvre has ended the results will be
shown at the bottom left of the screen of the screen.
To perform a second manoeuvre click Add Trial” and perform the second
test. When the second manoeuvre has been performed you will also be
able to see a % variation between the tests and also the variation in mls.
Do as many tests as required.
If after any manoeuvres you would like to delete them or
mark them as acceptable, double-click on the column of
that test.
A new window will appear. Press the “Delete” button to
delete that trial. A login prompt will appear. Type your
name into the box and click “login.
If the test was deemed to be unacceptable by the software
but you felt that the test was of good enough quality then
you may also override this by ticking Acceptable.
This will then allow that test to be used in the results.
When you have nished the required number of SVC tests
click on Test” (on the right side of the screen.
Then click on FVC (ex \only) on the main menu.
Conrm the temperature etc. and set the baseline by
blocking the Spirette (as per point 7) and then perform the
FVC tests as required.
Again after each test you will see the results appear at the
bottom of the screen and the variation between tests.
When you have nished the required number of FVC tests click on either
print to print or if you wish to save the test as a pdf click on report then
print menu” and then pdf and you can save it to the required folder.