3. DESCRIPTION OF MODES
Smulaon mode:
L1 - 5 kPa; L2 - 12 kPa; L3 - 17 kPa; L4 - 21 kPa ; L5 - 25 kPa ;
:Expression mode
L1 - 21 kPa; L2 - 24 kPa; L3 - 27 kPa; L4 - 31 kPa ; L5 - 37 kPa ; L6 - 40 kPa ;
L7 - 42 kPa; L8 - 43 kPa; L9 - 45 kPa;
Allowable difference is ± 5 kpa
4. DESCRIPTION OF ICONS AND BUTTONS
1. Power buon:
Hold for 2 seconds to turn the breast pump on or off.
2. Mode change:
Press the buon to change the mode. The device will automacally switch from smulaon to
expression mode aer 2 minutes. The device can save sengs and has a me counter. When you
turn on the device, the sengs from the previous use of the device are automacally turned on.
The breast pump will automacally turn off aer 30 minutes of use.
3. Sucon power:
+ increases sucon power; - reducon of sucon power (smulaon mode: L1-L5;
expression mode: L1-L9).
4. Baery status:
Indicates the amount of baery power remaining. Charge the breast pump if the baery is low.
5. Session duraon:
Indicates how long you have been using the breast pump. The device will turn off automacally aer
30 minutes of use.
E- Sucon increase
F- Sucon level
G- Expression mode
H- Session Duraon
See - Figure 2 on page 2
A- Sucon decrease
B- Smulaon mode
C- Baery Percent
D- Power buon/ mode change
5. BEFORE FIRST USE
6. CLEANING
Cleaning and sanizing are two different acvies. They must be done separately to protect
your child and ensure that the breast pump works properly.
Cleaning - washing the surface of the device and removing dirt.
Sanizing - killing organisms such as bacteria or viruses that may be present on the surface
of the device.
See - Figure 3 on page 2
1. Remove the cap from the breast shield before each use.
2. Pull down on valve toremove from connector.
3. Separate the connectorfrom the sucon bowl and sucon cap