ST Depression: Identical to ST Elevation except the ST segments are now depressed.
Selecting MI Myocardial Infarction branches to:
Myocard_In
f
Isc In
j
Inf Iin
f
Isc: Ischemia
Inj: Injury
Inf: Infarction
Iinf: Inferior Infarction
Isc: Ischemia : Normal Sinus Rhythm (NSR) with fully inverted T waves. This is a
condition of reduced blood supply to the heart in a normal patient.
INJ: Injury. A waveform with ST elevation of 20% with a negative slope and inverted T
wave.
Inf: Infarction. Normal Sinus Rhythm (NSR) with a large Q wave with the amplitude
increased six times and the width increased three times compared to normal.
Iinf: Inferior Infarction. A waveform with the Q wave modified as in Infarction and the ST
segment elevated 7% as in ST Elevation.
Selecting TalT branches to:
TalT/Rate = 80
Rate Am
l
An 80 BPM QRS test signal of 1 mV amplitude and 100 ms duration is generated with a T
wave duration of 180 ms and Q-T interval of 350 ms.
The T wave amplitude may be varied from 0 to 1.2 mV in steps of 0.1 mV .
Rate: 80 BPM
Amplitude: 0.0 0.1 0.2 0.3 0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2
Selecting RWD branches to:
RWD/
Widt Am
p
l
RWD: R Wave Detection. A QRS waveform is generated at 70 BPM with selectable width
and amplitude changes.
Widt: Width default setting 100 msec
10 40 50 60 70 80 90 100 110 120
Ampl: Amplitude default setting 1.0 mV
.15 0.3 0.5 1.0 2.0 3.0 4.0 5.0
Perf. Waves
SIN SQR TRI PLS
SIN: Sine Wave
SQR: Square Wave
TRI: Triangle Wave
PLS: Pulse Wave
SIN/
Fre
q
Am
p
l
Minisim 1000 User Manual 11