QRS Ventricular Tachycardia
Rate Amp Width Rate Amp
(BPM) (mV) (ms) (BPM) (mV)
80 1 100 195 2.0
80 1 100 195 1.0
80 1 100 195 4.0
RESPIRATION:
Select Resp in the Simulator Menu to access the Respiration selections.
Resp Menu
Rate Im
p
dR A
p
ne
Respiration waveforms are generated with four selectable rates, baseline impedances, and delta
impedance variations.
The Respiration default settings are:
Respiration Rate 30 BPM
Impedance 500 Ohms
Delta Impedance 1.0 Ohm
Apnea Off
The value changes that may be made are:
Rate: 15, 30, 60, 120 BPM.
Impedance: 250, 500, 750, and 1000 Ohms.
Delta Impedance: 0.1, 0.5, 1.0, and 1.5 Ohms.
Apnea:
Resp/Apne=Of
f
Off Cont 12s 32s
Off: Apnea is absent. Normal respiration waveforms are generated.
Cont: Continuous Apnea. No respiration waveforms are generated.
12s: No respiration waveform is generated for 12 seconds.
32s: No respiration waveform is generated for 32 seconds.
BLOOD PRESSURE:
Select Bp in the Simulator Menu to access Blood Pressure selections.
Two blood pressure waveforms are generated with selections of 12 static and 6 dynamic values.
BP1 values are selected from the listed choices and BP2 values are ½ of those selected for BP1.
BP1 Menu
D
y
na Stat
The Dynamic pressure waveforms track the Normal Sinus Rhythm rates.
The default settings for pressure values are:
BP1 Dynamic: 120/80 Static: 0
BP2 Dynamic: 60/40 Static: 0
The Dynamic pressure value selections are:
100/60 120/80
50/10 60/20
70/30 80/40
The Static pressure value selections are:
0 5 10 20
25 30 40 50
100 150 200 300
Minisim 1000 User Manual 16