Guidance and manufacturer’s declaration - electromagnetic
emissions - for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration - electromagnetic
immunity - for all EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic emissions
The Medical Infrared Thermometer is intended for use in the
electromagnetic environment specified below. The customer or user of
the device should ensure that it is only used in such an environment.
Emissions
test
Compliance
Electromagnetic environment - guidance
As this device only uses RF energy for its
internal functions, its RF emissions are very
low and not likely to cause any interference
to nearby electronic equipment.
The device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network which supplies buildings for
domestic purposes.
Group 1
Class B
RF emissions
CISPR 11
RF emissions
CISPR 11
Guidance and manufacturer’s declaration - electromagnetic immunity
The Medical Infrared Thermometer is intended for use in the
electromagnetic environment specified below. The customer or user of
the device should ensure that it is only used in such an environment.
Immunity test
IEC 60601
test level
Electromagnetic environment -
guidance
Floors should be wooden,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Compliance
level
±6KV
contact
±8KV air
±6KV
contact
±8KV air
3 A/m 3 A/m
Electrostatic
discharge
(ESD)
IEC
61000-4-2
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8