Re-insert the chamber and push it until you hear an audible click.
Warning
To avoid a burn injury, be careful not to touch the heater plate or the heated water in the water
chamber when the humidifier is switched on or has not yet cooled down. Wait 10 minutes for the
heater plate and water to cool.
Always stop treatment before detaching or attaching the water chamber.
Never add or pour out water from the water chamber when it is attached to the ventilator.
Caution
• Use only distilled or sterilised water or boiled, chilled tap water in the
humidifier water chamber. This is to reduce mineral deposits and maximise
the life of the water chamber.
• Do not fill the water chamber with hot water.
• Do not overfill the water chamber. Fill only to the maximum level indicated on
the water chamber.
• Always ensure the lid with seal is properly mounted after filling and
reassembling the water chamber.
• Also check that the water chamber is correctly docked in place and locked to
the ventilator.
• Make sure all parts are dry before the ventilator is connected to the mains and
put into operation.
• Change the water in your humidifier every day to prevent mould and bacteria
growth. Any remaining water in the morning should be disposed of.
• Do not put any chemicals or additives into the water as these may irritate your
airway or damage the water tank.
• Do not attend your NIV review appointments with water in the humidifier
chamber as this will damage your machine.
• If you are getting a lot of moisture or water inside your mask and tubing, the
humidifier is set too high. Adjust your humidification settings.
You can also remove the lid
and fill the water through the
top of the chamber. Always
make sure the lid of the water
chamber is totally sealed
before re-inserting.