Chapter 3 Assembling
46
5
Attach an epi fluorescence attachment (for the compact/DIA + EPI-FL system)
ttach the P2-EFLI Epi Fluorescence Attachment to the zooming body.
See "3.1 [7] Attach an epi fluorescence attachment. (for the DIA + EPI-FL system)
6
Attach a relay box.
ttach the relay box to the P2-FMDN Focus Mount.
The arm for attaching to the P2-FU Focus Unit and the
metal fitting for attaching to the P2-FMDN Focus Mount
are supplied with the relay box. Use the metal fitting to
attach the relay box to the focus mount.
(1) Align the relay box main body with the two screw
holes on the metal fitting as shown in the figure, and
attach it by tightening the M5 hexagon socket head
bolts supplied with the relay box using the hex drive
(nominal designation: 4).
(2) If cable clamps are installed in the focus mount,
remove them by loosening the fixing screws using a
Phillips driver. These screw holes are used for
attaching the relay box.
Align the two screw holes on the metal fitting
attached to the relay box with the screw holes on
the focus mount as shown in the figure. Attach it by
tightening the M4 8-mm pan head screws for cable
clamps using a Phillips driver.
(3) Connect the cable of the relay box main body to
RELAY of the P2-CTLB Control Box.
Attaching the metal fitting to the relay box
Attaching the relay box to the focus unit