Chapter 3 Assembling
47
7
Connect signal cables.
Connect the cables required for detecting the device status.
Connect the round connector on the left side of the zooming body with the "ZOOM" connector on the relay box using
the signal cable supplied with the SMZ18 Zooming Body.
Connect the signal cable from the P2-EFLI Epi Fluorescence Attachment with the "FL" connector on the relay box.
8
Attach a tube.
A
ttach the tube to be used to the zooming body.
The procedure is the same as the procedure for attaching
a tube to the fluorescence attachment or the
epi-fluorescence attachment. See “3.1, step 8 Attach a
tube.”
Attaching the tube by tightening the screw
9
Attach eyepieces.
A
ttach the eyepieces to the tube.
For details, see “3.1, step 9 Attach eyepieces.”
10
Attach an objective.
A
ttach the objective to the zooming body.
The procedure for attaching/removing the objective to/from the zooming body is the same as for attaching/removing
it to/from the intelligent nosepiece. See “3.1, step 10 Attach an objective.”
FL
Round connector