ตัวอย่างการเข้าทำรายการ Damping
1
2
3
Basic Setup
Bas Sys Trim CHK
PV Decimal
Total Decimal
Damping(s)
Lcd Rotate
01.0
Bas Sys Trim CHK
PV Decimal
Total Decimal
Damping(s)
Bas Sys Trim CHK
Max:99.9
Min: 0.1
Damping(s)
02.0
01.0
Damping time
Damping time on the meter display and output. Set range o 1-99 9S (unit is "seconds").
Set as follows:
In measurement mode, click
“C/CE” two times into
configuration menu
Modify value
Current value
Lcd Rotate
Steps Operation instructions
Interface show
Click “↓” to choice Damping(s).
Click “→” into “damping time” setup
menu, “→” and “↓” key to set
Damping times
4
5
6
Bas Sys Trim CHK
Damping(s)
02.0
01.0
OK:C/CE
CANCEL:
→
01.0
Bas Sys Trim CHK
PV Decimal
Total Decimal
Damping(s)
Click “C/CE” key to quit setup menu, LCD
show confirm menu
Lcd Rotate
Click “C/CE” key to confirm and return
configuration menu(click”→” give up
modify
Click “C/CE” two times from configuration menu to measurement
menu, also you can continue other operation.
ทิศทางการไหล
Forward(Fwd) คือทิศทางการไหล ในท่อเป็นไปตาม ลูกศรด้านข้าง Flowmeter
Reverse(Rev) คือการไหล ย้อน ทิศทาง ลูกศร
Bidirection (Bid) การไหลได้ทั้ง 2 ทิศทาง