Do you have a question about the Nova StatStrip and is the answer not in the manual?
Test Time | 6 seconds |
---|---|
Calibration | Factory calibrated |
Operating Humidity | 10-90% RH |
Sample Size | 1.2 microliter |
Operating Temperature | 15-40°C |
Memory Capacity | 400 tests |
Intended Use | Professional use |
Sample Type | Capillary, venous, arterial, and neonatal blood |
Display | LCD |
Connectivity | USB |
Hematocrit Range | 20 - 65% |
Log in to the meter and enter operator identification for QC testing.
Access the QC function and scan the glucose test strip barcode.
Scan the QC bottle, insert the strip, and apply the QC sample to the strip.
Review the QC test result and accept it if within the acceptable range.
Guidance on troubleshooting when QC test results fail, including meter usage suspension.
Log in to the meter and enter operator identification for patient testing.
Access the patient test function and scan the glucose test strip barcode.
Input patient identification or CAP number for proficiency testing.
Collect and apply blood sample to the test strip, ensuring proper technique.
Manage patient test results by accepting, rejecting, or reviewing.
Adhere to safety, verify ID, warm site, prepare meter, and select appropriate puncture site.
Use a lancet to puncture the skin at the selected site.
Apply blood from the heel or micro-collection pipette to the test strip.
Apply pressure to stop bleeding and dispose of used materials correctly.
Understand glucose ranges and protocols for critical low/high results.