User Manual - R.Test Evolution 4 EN V2019-12/06/2019 Page 7
Emissions from the power
terminals
(Emissions conducted)
(CISPR 11)
R.TEST4 is suitable for use in all settings, including
home health care environment and the environment
of a professional health care facility.
Harmonic current emission
(IEC61000-3-2)
Voltage variations, voltage
fluctuations and flicker
(IEC61000-3-3)
The R.Test.4 is intended for use in an electromagnetic environment as specified below. The clinician and
the user should ensure that the R.Test 4 is used in such an environment
Test level according
to IEC60601
Test level according to
IEC60601
Electromagnetic
environment / remarks
Electrostatic
discharge (ESD)
(IEC61000-4-2)
8 kV in contact
± 15 kV to the air
8 kV in contact
± 15 kV to the air
Home health care
environment and an
environment of a
professional health care
facility.
Fast electrical
transients in
bursts
(IEC61000-4-4)
Not applicable (no relation
to the public electricity
grid)
Home health care
environment and an
environment of a
professional health care
facility.
± 1 kV in Differential
mode
± 2 kV in common
mode
Not applicable (no relation
to the public electricity
grid)
Home health care
environment and an
environment of a
professional health care
facility.
Magnetic field at
assigned
industrial
frequency
(IEC61000-4-8)
Home health care
environment and an
environment of a
professional health care
facility.