User Manual - R.Test Evolution 4 EN V2019-12/06/2019 Page 8
Electromagnetic immunity
The R.Test.4 is intended for use in an electromagnetic environment as specified below. The clinician and
the user should ensure that the R.Test 4 is used in such an environment
Test level according
to IEC60601
Test level according to
IEC60601
Electromagnetic
environment / remarks
0% UT for 0.5 cycles
At 0 °, 45 °, 90 °, 135
°, 180 °, 225 °, 270 °
and 315 °
0% UT for 1 cycle
70% UT for 25 cycles
at 50 Hz
For 30 cycles at 60 Hz
Single phase: at 0 °
Not applicable (no relation
to the public electricity
grid)
Home health care
environment and an
environment of a
professional health care
facility.
0% UT;
for 250 cycles at 50 Hz
for 300 cycles at 60 Hz
Not applicable (no relation
to the public electricity
grid)
Home health care
environment and an
environment of a
professional health care
facility.
Electromagnetic Immunity, Portable Radio Frequency Equipment
Immunity test
Test level
according to
IEC60601
Test level
according to
IEC60601
Electromagnetic environment / remarks
WARNING: RF portable communication devices (including peripherals such as antenna cables
and external antennas) should not be used closer than 30 cm (12 inches) to any part of the
R.Test.4, including the cables specified by the manufacturer. Otherwise, the performance of these
devices may be impaired.
Electrostatic
discharge (ESD)
(IEC61000-4-2)
3 V/m
80 MHz at 2.7
GHz
80 % MA at 1
kHz
10 V/m
80 MHz at 2.7
GHz
80 % MA at 1
kHz
3 V/m
80 MHz at 2.7
GHz
80 % MA at 1
kHz
10 V/m
80 MHz at 2.7
GHz
80 % MA at 1
kHz
Home health care environment
Professional health care institution

Table of Contents