User Manual - R.Test Evolution 4 EN V2019-12/06/2019 Page 9
Proximity fields
issued by RF
wireless
communication
devices
(IEC 61000-4-3
interim method)
9 V/m
710 MHz, 745
MHz,
780 MHZ,
5240 MHz,
5550 MHz,
5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810
MHz,
870 MHz, 930
MHz,
1720 MHz,
1845 MHz,
1970 MHz,
2450 MHz
9 V/m
710 MHz, 745
MHz,
780 MHZ,
5240 MHz,
5550 MHz,
5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810
MHz,
870 MHz, 930
MHz,
1720 MHz,
1845 MHz,
1970 MHz,
2450 MHz
Home health care environment and an
environment of a professional health care
facility.
Conducted
disturbances,
induced by RF
fields
(IEC610004-6)
3V 150KHz to
80MHz
6V in ISM
band and
band between
0.15 MHZ and
80 MHZ,
amateur radio
band included
80% MA at 1
KHz
3V 150KHz to
80MHz
6V in ISM
band and
band between
0.15 MHZ and
80 MHZ,
amateur radio
band included
80% MA at 1
KHz
Home health care environment and an
environment of a professional health care
facility.
The electromagnetic field strengths of fixed radio frequency transmitters, as determined by an
electromagnetic environment measurement (a), shall be less than the compliance level for each
frequency range. Interference may occur near equipment identified by the following symbol:
Note: These specifications may not apply in all situations. Electromagnetic propagation is affected by
the absorption and reflection of structures, objects and people.
(a) The intensities of the electromagnetic fields of fixed radio frequency transmitters, such as base
stations for mobile (cellular / wireless) telephones, mobile radios, radio amateurs, AM / FM radio
transmissions and TV transmissions cannot be accurately determined theoretically. To evaluate the
electromagnetic environment due to fixed radio frequency transmitters, an electromagnetic
environment measurement must be performed. If the measured RF field strength in the immediate
environment of use of the product exceeds the radio frequency compliance level (specified above), it
is necessary to test the product performance to verify that it conforms to the specifications. If