ALARM REASON SOLUTION
The set is not fitted or
is fitted incorrectly
Reinsert the giving set as per instructions
The pressure sensor
is dirty
Clean the sensors and reinsert cassette
Door is broken, cracked
or not closed
Make sure the door is closed
Tape door to pump tightly to continue
feeding and call your Nutricia Homeward
Customer Service team to arrange for a
replacement
Flow rate has not been
programmed
Make sure the correct flow rate is
programmed
The pump has been
paused for three or
more minutes
By pressing the START/STOP button
to silence the alarm, a further three
minute pause commences. When ready
to restart feeding press the START/
STOP button
Air bubble in the giving
set
Disconnect the giving set from the
feeding tube and hold over a clean
container. Press the FILL SET button until
the air bubble has passed out of the giving
set. Press FILL SET to stop, reattach to
feeding tube and press START
The bag of feed is empty
If further feeding is required, replace
the empty bag of feed and recommence
feeding. If you need to rell the
giving set with feed, follow solution
instructions for “air bubble in the giving
set”. If required dose delivered, switch
off pump and discard feed and giving
set as per training
The air sensor is dirty
Clean the sensor area following the
cleaning guidelines
ALARM REASON SOLUTION
The pump has
administered the dose
Press and hold the ON/OFF button
until it beeps and is powered off
If the END OF DOSE alarms
inappropriately contact your
Nutricia Homeward Enteral Nurse
Specialist or the Out of Hours
Advice Line
The pump detected an
occlusion between the
feed and the pump
Make sure there are no kinks in the
tubing, that clamps are not closed
between the feed container and the
pump, and that the foil has been
pierced
The sensor is dirty
Clean the sensor area (see cleaning
guidelines), reinsert the giving set in
the pump and restart the pump
The pump detected an
occlusion between the
pump and the patient
If relevant, make sure the clamp is open
Make sure there are no kinks in the
giving set or feeding tube
The sensor is dirty
Clean the sensor area (see cleaning
guidelines), reinsert the giving set in
the pump and restart the pump
Please refer to the step by step guide on next page è
CAUTION: Do not use the ‘FILL SET’ funcon when there is a set occlusion
as this may cause the pump to incorrectly calibrate
OCC OUT
OCC IN
DOOR
PUSH
START
AIR
NO SET
END OF
DOSE
HOW TO STOP THE OCC ALARM
Press the START/STOP button
Detach the giving set
Solve the occlusion problem (see below)
Re-insert the giving set and close door
Press the START/STOP button
THE MOST COMMON SET OCCLUSION
TUBE KINKING: Narrow the tube, release kinking
PROG

Other manuals for Nutricia Flocare Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nutricia Flocare Infinity and is the answer not in the manual?

Nutricia Flocare Infinity Specifications

General IconGeneral
Device TypeEnteral Feeding Pump
ManufacturerNutricia
ModelFlocare Infinity
Accuracy±5%
Flow Rate Increment1 ml/h
Keep Vein Open (KVO)1 ml/h
Power SupplyRechargeable battery or mains power
Bolus Dose1 to 400 ml
Bolus Interval1 to 24 hours
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Empty, Low Battery, Error
Storage Temperature-20 to 60°C
Battery Life24 hours
Operating Temperature5 to 40°C