EasyManua.ls Logo

Oakton EC1 - Self-Diagnosis Functions; Factory Calibration and Accuracy Check; Parameter Setting; Parameter Setting Procedure

Oakton EC1
8 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.2 The tester adopts 1413 µS and 12.88 mS standard calibration solutions. Users
can use 1-2 point calibrations as needed. For most circumstances, calibrating in
1413 µS to complete 1
st
point calibration will meet testing requirements.
5.3 Unit Information: 1 mS/cm = 1000 µS/cm
Once the readings are greater than 2000µS/cm, the unit will automatically
become mS/cm.
5.3 The tester has self-diagnosis functions:
Symbol
Self-Diagnosis information
How to fix
Wrong calibration solution,
which exceeds the
recognizable range of the
meter.
1. Check if calibration solution is
correct
2. Check if probe is damaged.
is pushed before
measurement is stable
( comes up and stays)
Wait for the smile icon to stay,
and then short press
5.4 The tester has already been calibrated after manufacture. Usually users can
use the tester right away, or test it in the calibration solutions to test its
accuracy. When error is large, calibrate it before using.
6. Parameter Setting
6.1 Parameter setting reference chart:
Symbol
Code
Factory Default
P1
Aut-µS-mS
Aut
P2
No Yes
No
6.2 How to setup parameters:
When turned off, long press to enter setup short press to switch
P1-P2 Short press , parameter flickeringshort press to choose,

Related product manuals