short press to confirm parameter selectionLong press to switch off.
6.3 Parameter setting notes
a) Select conductivity ranges (P1):
Aut-- All ranges: 0 199.9 µS/cm, 20 1999 µS/cm, 2.0 20.0 mS/cm;
Factory default is Aut.
b) Restore to factory default:
Select Yes to restore the calibration to the theoretical values and parameter
setting to original values. When meter’s calibration or measurement performs
abnormally, this function can be adopted so the meter goes back to factory
default setting and then users can conduct calibration or take measurements
again.
7. Technical Specifications
Conductivity
Range
0 200.0 µS/cm, 0 2000 µS/cm, 0
20.00 mS/cm
Resolution
0.1/1 µS/cm, 0.01 mS/cm
Accuracy
±1% F.S±1 digit
Calibration points
1-2 points auto calibration
Automatic Temp.
Compensation
0-50˚C
Temp.
Range
0-50˚C
Resolution
0.1˚C
Accuracy
±0.5˚C
8. Other Functions & Parameters