VIII. Subsidiary Functions with the Meter (for reference)
1. Averaging: when the meter measures the same sample twice or more, press the “确定”(OK) key once,
and the average value of previous measurements will be shown, such as
.
Note 平均Average
2. Indicating sample weight: After the meter displays the first measured moisture value or the average
value, press the “确定”(OK) key to display the sample weight in gram; and press the “确定”(OK) key
to redisplay the moisture value.
3. Indicating sample temperature: under the status of weight indication, press the “Category” key to
display the sample temperature in ; and press the “确定”(OK) key again to redisplay the moisture
value.
4. Volume weight conversion: the volume of the blanking cylinder supplied with the meter is about
232cm
3
(ml) without using the funnel. The sample shall be stickled with the funnel along the edge of
blanking cylinder before being put into the meter for measurement. After showing the moisture value,
press the “确定”(OK) key to display the sample weight. Then press the “确定”(OK) key again to starts
automatic conversion, shortly after which the volume weight of the sample will be displayed, e.g.,
. This value, of course, is for reference only.
5. Communication interface: used for special occasions. Please contact us for its technical parameters
and usage.
IX. Accessories of the Meter
The instrument case shall include the following accessories:
AC adapter, cleaning brush, user’s manual, quality conformity certificate and quality warranty card,
blanking cylinder, funnel, calibration weight for each meter; four 5# alkaline batteries.
X. Major Technical Data
Measuring object: foodstuff and other non-metal granular samples, such as rice grain, wheat, corn,
soybean and rapeseed;
Measuring error: ±0.5% Repetitive error: 0.2%
Measuring range: 3~35% Measuring time: 10s
Environmental temperature: 0~40 Net weight: 830g
Power supply: four 5# alkaline batteries or external 9V DC stabilized power source
Display mode: high-brightness backlight LCD