MNL-1029 Rev A
16
Symbol
Description
Time
Notes
min
typ
max
T0
Software Trigger Command
Start of Acquisition Cycle
t
PROC
Trigger Command Processing
System dependent (e.g., computer
hardware, software, driver)
t
ACQDLY
Acquisition Delay
0 µs
--
21.47 s
User specified, 1 µs resolution
t
BUSY1
Busy
1 µs
--
218 µs
1 µs for t
INTEG
>= t
BUSY2
(219 - t
INTEG
) µs for t
INTEG
< t
BUSY2
t
INTEG
Integration Period
1 µs
--
1.19 hours
User specified, 1 µs resolution
t
BUSY2
Busy
--
--
--
Constant at 218 µs
t
SSDLY
Single Strobe Delay
0 µs
--
1.19 hours
User specified, 1 µs resolution
t
SSH
Single Strobe High Time
0 µs
--
1.19 hours
User specified, 1 µs resolution
t
CSPER
Continuous Strobe Period
0 µs
--
1.19 hours
User specified, 1 µs resolution
t
CSOFF
Continuous Strobe Off Time
0 µs
--
t
CSPER
The following timing diagram represents the acquisition process for a Software Trigger event.